SRP
Rules & Regulations Guidance
Documents AGWS
Previous
Table of Contents
AGWST 6.00
Alternative Ground Water
Sampling Techniques Guide (July 1994)
Title: Ground Water Sampling with the
use of a HydroPunchR Direct Push Sampler (3/94)
Method Number: AGWST 6.00
Summary:
The HydroPunchR is a sampling tool constructed
of stainless steel and teflon used for collecting ground
water samples. This document provides guidance for the use
of this tool in ground water investigations.
I PURPOSE AND SCOPE
This document summarizes the minimum requirements
for the use of the HydroPunch® (HP-I and
HP-II) for the collection of ground water data for site
investigations.
II METHOD OVERVIEW
-
Tool
-
The HydroPunch® I (HP-I) sampling
tool collects the sample in only one mode, within
the sample chamber (Figure
I). This tool collects ground water through
the effect of in-situ hydrostatic head, therefore,
the top of the sample chamber must be below the
ground water table for sample acquisition. A sample
cannot be collected across the ground water table
with the HP-I. The HP-I is designed to be used by
cone penetrometer or drill rig.
-
The HydroPunch® II (HP-II) sampling
tool can be operated in two modes, hydrocarbon and
water sampling (Figure
II). The water sampling mode is similar in operation
to the HP-I. In the hydrocarbon mode a PVC screen
is exposed so samples can be collected across the
ground water table of an unconfined aquifer to determine
the presence of floating product. The HP-II was
specifically designed to be used by drilling contractors.
Its larger diameter limits the effective depth when
pushed from the surface with cone penetrometer rigs.
-
Application
-
Collection of ground water samples for the determination
of the presence/absence and extent of ground water
contamination.
-
Field screening tool to aid in the placement of
monitor wells.
-
Temporary placement for the collection of ground
water samples and estimating ground water flow directions
(less than 48 hours).
-
Capabilities
-
Obtain ground water samples from unconfined aquifers.
-
Obtain ground water samples from confined aquifers
provided the upper aquifer is cased off and the
casing is driven a minimum of two feet into the
confining layer.
-
Obtain samples across the water table to determine
the presence of floating product (HP-II).
-
Capable of collecting samples to determine the
vertical profiling of contaminants in an aquifer.
-
Ability to collect ground water samples from small
discrete water bearing zones. (HP-I & HP-II)
-
Capable of being used with a cone penetrometer
rig or a conventional drill rig.
-
7. A comparison of the advantages and limitations
for both the HP-1 and HP-II are listed in Table
I.
III SAMPLING METHOD REQUIREMENTS
-
Installation
-
The HydroPunch® is capable of use
in unconsolidated formations only. When being installed,
the drilling must stop above the target sample depth
thereby not disturbing the zone to be sampled. It
is therefore imperative to have some idea of the
depth at which the sample will be collected. If
little is known of the site geology, then an initial
boring should be made to determine 1) depth of water
bearing zones 2) permeability of sample zone 3)
density of soil 4) identify the subsurface stratigraphy
5) other pertinent data for the investigation.
-
When used with a conventional drill rig the hole
must be advanced (with hollow stem augers, mud rotary
etc.) to the depth which is above the zone of interest,
eliminating any interference from the drilling.
The HydroPunchR may then be driven to the desired
sampling interval for sample collection.
-
DO NOT set the HydroPunch® down
on the bottom of the borehole and pick it up. This
will open the tool and compromise the sample integrity.
Damage to the tool may be incurred if it is driven
after being opened. Also, caution must be taken
not to back hammer when driving the HydroPunchR
for the above stated reason.
-
Always accurately measure the distance the tool
is pushed or driven and the distance pulled back.
-
Never pull the HydroPunch® back
farther than it is pushed or driven into the undisturbed
soil. This may result in cross contamination of
the sample from other zones in the borehole, or
loss of the casing (in the hydrocarbon mode) resulting
in the inability for sample collection.
-
Installation of the tool is required to comply
with all permit, license, sealing and grouting requirements
as per Appendices I and
II. Any tool left in the
ground longer than 48 hours is considered a monitor
well and therefore must comply the permit, installation
and license requirements for monitor wells.
-
Sampling Procedures
-
Hydrocarbon Mode (HP-II)
-
The hydrocarbon mode is used to collect ground
water samples when:
-
A sample must be obtained from the water
table interface of an unconfined aquifer.
-
A large volume of sample is required.
-
The presence of floating product is suspected.
-
A sacrificial 0.010-inch PVC screen (approx.
5') is attached to a disposable drive cone.
The screen and drive cone are then inserted
into the body of the HP-II until the O-Ring
on the cone is sealed in drive shoe. Place the
sleeve over the juncture of the drive cone and
body of the unit.
-
Once driven to the desired depth, the body
of the unit is pulled back exposing the screen.
Friction with the seal will hold the cone in
position while the screen is exposed. Do not
pull back a distance greater than the length
of the screen.
-
The EW, BW, or NW casing used to drive the
tool allows for the placement of a small diameter
bailer (3/4" or 1") to be lowered down through
the casing and body of the HP-II and into the
screen for sample collection.
-
The Hydropunch does not have to be purged or
developed prior to sampling.
-
Water Sampling Mode (HP-1 and HP-II)
-
The HP-II in the ground water sampling mode
or HP-I can be used when samples are required
at a minimum of five (5) feet below the top
of the water table and when a small sample volume
(500 ml-1,200 ml dependent upon tool) is adequate
-
Place the lower check valve with attached
filter screen into the bottom of the tool body
and place the upper check valve in the top of
the tool. Insert the disposable drive cone into
the drive shoe ensuring a seal is made by the
O-Ring. Place sleeve over the juncture of the
drive cone and drive shoe.
-
Push or drive (with 140 lb hammer, 30 inch
travel) the unit to the desired depth and pull
back approximately two (2) feet. Soil friction
will hold the drive cone in place.
-
Ground water flows into the intake screen
past the lower check valve, into the sample
chamber and finally out the top check valve.
-
When full the tool is pulled to the surface,
increasing the hydrostatic head within the tool
closing the two check valves.
-
At the surface the HP-II is inverted and the
sample is decanted through a discharge valve
and tubing into the sample containers.
-
Quality Assurance/Quality Control
-
Decontamination
The HydroPunch®, drill rods and drive
casing must be decontaminated between samples using
the following procedure:
-
Disassemble the HydroPunch® unit
and remove O-Rings.
The PVC screen is disposable and must be discarded.
-
Scrub with a laboratory grade glassware detergent.
-
Rinse with potable water and/or steam clean.
-
Rinse entire unit with distilled and deionized
ASTM Type II water.
-
Replace O-Rings.
-
Reassemble unit.
-
The PVC screen is supplied by the manufacturer
already cleaned. If the packaging is compromised
then it should be cleaned in the same manner as
the HydroPunch and casing.
-
Field Blanks
Field blanks must be obtained in the same manner
as samples (i.e., if hydrocarbon mode is used blank
water must pass through bailer, screen and HydroPunch®
body).
Parameters and frequency for field blanks are designated
in the May 1992 edition of the NJDEPE Field Sampling
Procedures Manual (FSPM).
-
Sample Equipment
The NJDEPE Field Sampling Procedures Manual can be
used as a reference for the selection of sampling
equipment and procedures for use with the HP-II in
the hydrocarbon mode. The HydroPunchR in the water
sampling mode is in itself a sampler.
All sampling equipment must be decontaminated in
accordance with the NJDEPE Field Sampling Procedures
Manual and dedicated to each sample point.
-
Rod Sealing
When using the HydroPunchR in the hydrocarbon or
ground water mode for obtaining samples deep in the
unconfined aquifer or in a confined aquifer, or using
the unit with hollow stem augers on mud rotary drilling,
the drill rod/casing joints must be sealed. This will
prevent fluid from entering the rods and potentially
contaminating the sample. The rods should be sealed
with TeflonR tape on the threads. Once put together
the joints must be sealed with gas pipe tape. Another
option is the use of drill rod with O-Rings at the
threads for sealing.
-
Formation Types
The HydroPunch® can be installed in
unconsolidated materials. Varying amounts of pebbles,
cobbles and boulders may impede advancement or damage
the tool.
IV REFERENCES
-
Cordry, Kent; "Hydropunch ® User's
Guide
-
Cordry, Kent; "Technical Information and Application
Guidelines - Hydropunch"
-
Bergen, C.L.; Tuckfield, R.C.; Park, NM; "Suitability
of the Hydropunch for Assessing Ground Water Contaminated
by Volatile Organics
-
Cordry, Kent; HydroPunch II - The Second Generation.
A New In Situ Ground Water Sampling Tool. In Procedures
of the Fifth National Outdoor Action Conference on Aquifer
Restoration Ground Water Monitoring, and Geophysical
Methods. pp 715-723 May 13-16, 1991, Las Vegas, Nevada.
-
Strutynsky, A.I.; Sainey, T.J.; Use of Piezometeric
Cone Penetration Testing and Penetrometer Ground Water
Sampling for Volatile Organic Contaminant Plume Detection.
In Procedures of the Petroleum Hydrocarbons and Organic
Chemicals in Ground Water: Prevention, Detection and
Restoration. p70-84. October 1990, Houston, TX.
-
Edge, R; Cordry, K; The HydroPunch: An In Situ Sampling
Tool for Collecting Ground Water from Unconsolidated
Sediments. Ground Water Monitoring Review, Vol. IX (3)
pp 177-183, 1989.
-
Smolley, M; Kappmeyer, J; Cone Penetrometer Tests
and HydroPunch® Sampling: A Screening
Technique for Plume Definition. Ground Water Monitoring
Review, Vol XI, No. 3, pp 101-106.
-
Van Sciver, C., Wallace, E.; The Evaluation of the
HydroPunch II to Obtain a Representative Ground Water
Sample. 9th Annual Waste Testing and Quality Assurance
Symposium; July 12-16, 1993, Arlington, VA
TABLE
I
ADVANTAGES AND LIMITATIONS
COMPARISON OF HP-I AND HP-II
|
|
HP-I |
|
HP-II |
ADVANTAGES |
|
-
Small diameter - can be used with cone
penetrometer rig.
-
Reusable cone.
-
Vertical profiling from a single borehole
without concern about drilling through disposable
cones and screens.
|
|
General:
-
Simpler design and fewer parts for fast
decontamination.
-
No moving parts are attached permanently
to the tool making it more durable and reliable.
-
Removable check valves providing 2 sample
modes which increases flexibility.
Hydrocarbon Mode:
-
Can collect sample at top of aquifer,
including product.
-
Can collect an unlimited volume of sample.
-
Can collect sample from thin aquifer.
-
Can directly measure fill rate.
Ground Water Mode:
-
Tool does not have to be driven on special
casing.
-
Only tool needs to be decontaminated.
-
Tool can be driven using downhole wireline
hammers.
|
LIMITATIONS |
|
-
Thin diameter and sliding parts with close
tolerances make tool susceptible to damage
when driven by drilling rig.
-
Short intake interval (11- inch) makes
sampling from thin water bearing zones difficult.
-
The intake screen must be at least 5 feet
below the top of the aquifer to collect
a complete sample.
-
Sample volume is limited to approximately
500 ml.
-
Yields a turbid sample.
-
Sample time intervals in low aquifers
may cause degeneration of sample integrity.
-
Requires drill rig for installation.
|
|
Hydrocarbon Mode:
-
Hollow drive pipe must extend to surface.
-
Drive pipe must be decontaminated.
-
A cone and screen is lost each time the
tool is used.
-
The rate and amount of sample obtained
is dependent upon the permeability of the
formation.
-
Yields a turbid sample therefore samples
for various analytes may be biased high.
-
Long sample acquisition times in low yielding
aquifers may cause degeneration of sample
integrity.
-
Requires drill rig for installation.
Ground Water Mode:
-
The intake must be at least 5 feet below
the top of the aquifer to obtain a full
sample.
-
Direct monitoring of the tool fill rate
is difficult.
-
Sample volume is limited to 1.2 liters.
-
Yields a turbid sample therefore samples
for various analytes may be biased high.
-
Long sample acquisition times in low yielding
aquifers may cause degeneration of sample
integrity.
-
Requires drill rig for installation.
-
Formations with 20-30% silts and clays
may not yield sufficient water for sampling
and limit use of the tool
|
|
Next
|
 |